A Ferromagnetic Surgical System Reduces Phrenic Nerve Injury in Redo Congenital Cardiac Surgery
Authors: Takeshi Shinkawa, Jessica Holloway, Xinyu Tang, Jeffrey M. Gossett, Michiaki Imamura
First Published: January 31, 2017
Abstract
A ferromagnetic surgical system (FMwand®) is a new type of dissection device expected to reduce the risk of adjacent tissue damage. We reviewed 426 congenital cardiac operations with cardiopulmonary bypass through redo sternotomy to assess if this device prevented phrenic nerve injury. The ferromagnetic surgical system was used in 203 operations (47.7%) with regular electrocautery and scissors. The preoperative and operative details were similar between the operations with or without the ferromagnetic surgical system. The incidence of phrenic nerve injury was significantly lower with the ferromagnetic surgical system (0% vs 2.7%, P = 0.031). A logistic regression model showed that the use of the ferromagnetic surgical system was significantly associated with reduced odds of phrenic nerve injury (P < 0.001).